5.4
Continuity Test 
➭ Disconnect supply power from the electrical circuit of the 
device to be measured, and discharge all high-voltage 
capacitors.
➭ Make sure that the device under test is voltage-free. 
Interference voltages distort measurement results!
➭ Set the rotary switch to “
”.
➭ A loudspeaker symbol appears at the display.
➭ Connect the conductor path under test as shown.
Depending upon the selected limit value, the multimeter 
generates a continuous acoustic signal in the case of continuity or 
short-circuiting, i.e. at a value of less than the selected limit value.
0L” appears at the display in the case of an open connection.
The limit value can be adjusted in the “
SET” menu (see also 
 
1nFo 
Z
 ... 
Z
 
SET 
 
rAtE 
V
 ... 
V
 
bEEP 
 
1, 10, 20, 30, 40, 90 
Ω 
UV
 
 
(10 = default setting)
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Temp
Ω MΩ
ISO
0 V  !
!
0.443
V
DC
R < 
1, 10, 20, 30, 40, 90 
Ω
FUNC
ENTER
R
:  300
Ω
Measuring Range:
000.8
Ω
(0.1 
Ω to 310 Ω)