Gossen Metrawatt Metrahit Iso Insulation measuring device, CAT III 600 V Metrahit Iso Benutzerhandbuch

Produktcode
Metrahit Iso
Seite von 72
38
GMC-I Gossen-Metrawatt GmbH
Measurements: V/Hz, 
Ω,  , Temp (°C/°F), 
, M
Ω
@UISO
, A/Hz
5.6
Capacitance Measurement 
 
➭ Disconnect supply power from the electrical circuit of the device 
to be measured, and discharge all high-voltage capacitors.
➭ Make sure that the device under test is voltage-free. 
Capacitors must always be discharged before measurement is 
performed. 
Interference voltages distort measurement results! 
Refer to section 5.1.1 regarding testing for the absence of 
voltage with the help of the direct voltage measurement.
➭ Set the rotary switch to “
”.
➭ Connect the (discharged!) device under test to the sockets 
with the measurement cables as shown.
Note

The “–” pole of polarized capacitors must be connected to the 
⊥” jack. 
Resistors and semiconductor paths connected in parallel to 
the capacitor distort measurement results!
COM mA A
V
Temp
Ω MΩ
ISO
COM mA A
V
Temp
Ω MΩ
ISO
0 V  !
!
10.38
F
+
+
μ
Measuring range: 10 pF to 300 
μF
5 ranges: 30 nF, 300 nF, 
μF, 30 μF and 300 μF