11.3a 
Touch current (substitute leakage current)
The limit value is:
0.5 mA
This measurement must also be carried out on devices of
protection class I with touchable conductive parts that are not
connected to the protective conductor.
Press the [OK] button to go to the ‘Touch current’ measuring display.
The measurement is made in accordance with the substitute leakage
current principle.
Here, too, the corresponding limit value is displayed beside the measured
value (Fig. 39).
The star sign on the left side of the display blinks when the
measuring process is running (the values are being measured).
The 
[OK]
 button appears when the measured value is determined.
A help display is also available if you press the [Help] button.
Warning!
With the test procedure according to DIN VDE 0702 and a test with
insulation resistance
, the touch current is always measured
by the tester with connection to the mains voltage (direct method
or according to the differential current principle).
11.3b 
Touch current (compare with Point 11.2)
The limit value is:
0.5 mA
This measurement must also be carried out on devices of
protection class I with touchable conductive parts that are not
connected to the protective conductor.
If you have selected a test procedure without insulation resistance in the
‘Settings for measuring’, the tester changes to measuring the touch current
(Points 11.2. Insulation resistance and 11.3a. Touch current / Substitute
leakage current, are omitted)
Caution! The test object is supplied with mains voltage
during the test.
After pressing the [->Next] button the display changes and you receive a
warning message about the connection to the mains current (Fig. 40).
With your assent – by pressing the large button – the display changes to
measuring the touch current and the mains voltage is connected – the test
object is put into operation!
You can see the measured value in the middle of the display, with the
related limit value to the right (Fig. 41).