3. Safety instructions
The appliance tester TG uni 1 was constructed and tested in accordance
with the applicable safety regulations and left the workshop in a faultless
safety condition. In order to maintain this condition and ensure safe
operation, the user must observe the instructions and warnings contained in
this operating manual.
With all work, the applicable accidend prevention regulations of the
trade associations for electrical systems and equipment must be
observed.
To avoid electric shock, the applicable safety regulations 
and DIN-
VDE regulations regarding high touch voltage 
must be observed
without fail when working with voltages greater than 120V (60 V)
DC or 50 V (25 V) eff. AC. The values in brackets apply for
restricted areas (e.g. medicine, agriculture).
Measuring at a dangerous proximity to electrical systems should only be
carried out under the instruction of a responsible electrician, and never
alone.
Check the tester and the connecting cables for external damage before
every new operation.
Make sure that the tester and the connecting cables are in faultless
condition. The tester may not be used if one or more functions fail or if
functional readiness is not evident.
The measuring lines and the measuring accessories may only be
touched in the designated handling areas.
Touching measuring connections, e.g. test tips must be avoided
under all circumstances.
If the safety of the operator is no longer guaranteed, the tester must
be decommissioned and secured against unwanted use. This is
the case if the device:
– shows obvious signs of damage
– no longer carries out the required measurements
– has been stored in adverse conditions for too long
The tester may only be used in the operating and measuring ranges
specified under Technical Data.
Prevent overheating of the tester by exposure to direct sunlight.
Only this will guarantee faultless functioning and a long sevice life of the
device.
Never open the housing of the tester because of the danger of high
voltages. The tester contains no parts that can be replaced by the user.
Only let the tester be serviced by qualified personnel.
6
Operating manual for the TG uni 1 appliance tester
Safety instructions
&
&
&
&
&
&
&
&
&
&