Rgb Lasersysteme QWAVE UV Lux-Meter, illumination measuring device, Brightness meter, DSA491006 Datenbogen

Produktcode
DSA491006
Seite von 30
Operation 
 
13 
To  take  a  reference  spectrum  for  transmission  or  absorption  spectroscopy,  turn  on  the 
light  source  again,  remove  the  sample,  set  an  appropriate  exposure  time  and  temporal 
averaging  and  click  the  “Take  Reference  Spectrum”  button  [ ].  After  selecting  the  “Use 
Reference  Spectrum”  button  [ ],  you  can  take  spectra  of  your  sample  for  measuring 
absorption or transmission. 
To  take  a  reference  spectrum  for  reflection  spectroscopy,  you  need  a  white  reference 
sample  to  calibrate  the  setup  for  100  %  reflection.  However,  all  real-world  reference 
samples  have  a  reflection  of  less  than  100  %  and  an  uneven  spectral  distribution. 
Therefore, for accurate measurements you also need to supply a reference spectrum for 
your  reference  sample.  To  take  a  reference  spectrum,  place  the  white  reference  sample 
where  you’ll  later  place  the  sample  to  be  measured.  Then  set  an  appropriate  exposure 
time and temporal averaging and click the “Take Reference Spectrum” button [ ]. In the 
following  dialog  window,  you  can  also  specify  a  data  file  that  includes  a  calibrated 
spectrum of your reference sample. After selecting the “Use Reference Spectrum” button 
[ ], you can measure reflectivity spectra. 
The  y  axis  of  the  spectrum  now  shows  the  relative  intensity  of  the  measured  spectrum 
compared to the reference spectrum. In the “Options” dialog from the File menu you can 
choose different units for displaying the spectra: 
•  Ratio  shows  the  ratio  of  the  measured  spectrum  to  the  reference  spectrum  as  in 
the formula for ! λ  above. 
•  Percentage shows the same value ! λ , but written as a percentage (i.e. multiplied 
by 100).
2
 This unit is usually used for transmittance. 
•  Absorbance shows the spectral absorbance ! λ  according to the formula above. 
You can choose between natural and base 10 logarithm. 
These  options  only  determine  how  to  display  the  spectra  in  the  diagram.  The  do  not 
modify the actual spectrum data. 
If  your  light  source  does  not  cover  the  full  range  of  the  spectrometer,  the  calculated 
spectrum  may  show  large  amounts  of  noise  outside  of  the  wavelength  range  of  your 
light  source.  This  is  normal  and  reflects  the  fact  that  the  measurement  has  a  large 
uncertainty at these wavelengths. 
These  measurements  are  quite  sensitive  to  changes  of  the  sample  illumination.  If  in 
doubt, you may want to take another reference spectrum to check if it is still valid. 
3.9  Calibration 
The  calibration  for  the  wavelengths,  dark  spectra  and  spectral  sensitivity  are  stored  on 
the device and loaded automatically during device initialization. So in contrast to almost 
any other spectroscopy software, you don’t need to worry about taking dark spectra and 
other calibration parameters, it’s all done automatically. 
                                                                              
2
  Please note that „percent“ means nothing else than “divided by 100”.