Rgb Lasersysteme QWAVE UV Lux-Meter, illumination measuring device, Brightness meter, DSA491006 Datenbogen

Produktcode
DSA491006
Seite von 30
Operation 
 
15 
combo box select the type of light source used. If you don’t find your light source in the 
list, you need to supply your own calibration lines. In this case, please take a look at the 
“CalibrationLines  xxx.dat”  files  in  the  installation  folder  on  your  hard  drive  and 
create a new similar file containing the spectrum lines of your light source. 
After you have selected your light source, click on “Start Fit”. The software tries to match 
the peaks in the measured spectrum and the given spectrum lines for your light source 
automatically.  Not  all  of  the  given  spectrum  lines  have  to  be  present  in  the  measured 
spectrum,  but  all  measured  peaks  must  be  known  spectrum  lines.  If  a  match  has  been 
found,  the  average  deviation  is  displayed.  If  the  match  seems  to  be  successful,  it  is 
displayed  in  green,  otherwise  in  red.  You  can  also  graphically  display  the  deviations  by 
clicking  on  “Show  Deviations”.  If  the  measured  peaks  have  successfully  been  identified, 
you  can  set  the  four  new  wavelength  calibration  parameters  by  clicking  on  the  “Save” 
button and confirm by clicking “OK”. The new wavelength calibration is then saved to the 
device. 
Dark Spectra 
Inside the Qwave spectrometer, the spectra are detected with a CCD line sensor. Like all 
image  sensors,  this  sensor  detects  a  small  signal  even  if  there  is  no  light  entering  the 
spectrometer.  This  signal  depends  on  the  individual  pixel,  the  exposure  time  and  the 
temperature. In order to obtain a precise measurement with a small noise level, this “Dark 
Spectrum”  must  be  taken  while  no  light  enters  the  spectrometer  and  then  subtracted 
from the actual measurement. 
For your Qwave spectrometer, this dark spectrum has already been taken, is stored in the 
device and automatically subtracted from the measured spectrum each time you take a 
new spectrum. Because the dark spectrum depends on the exposure time in a non-linear 
way,  there are actually multiple dark spectra stored in the device for a certain number of 
exposure  times.  The  dark  spectrum  that  is  needed  for  a  chosen  exposure  time  is  then  
obtained by interpolating between these dark spectra. 
In  case  you  find  that  the  spectra  you  get  are  not  zero  on  the  average,  even  if  no  light 
enters the spectrometer, then you probably need to recalibrate the dark spectra. You can 
also  check  the  dark  spectrum  calibration  using  the  “Calibrate  Dark  Spectrum”  window 
(see below). 
To recalibrate the dark spectra, simply close the spectrometer entrance, make sure that 
no  light  enters  the  device  and  choose  “Calibrate  Dark  Spectrum”  from  the  Tools  menu. 
Then  click  the  “Take”  button  and  go  for  a  coffee.  After  a  few  minutes  the  software  has 
taken dark spectra for a series of exposure times. You can check the new dark spectra by 
clicking on the “Test Button”.  While the test is running, you can in the lower diagram the 
spectra  taken  without  (red)  and  with  (green)  dark  spectrum  correction.  In  the  upper 
diagram you can see the noise level, also without (red) and with (green) dark spectrum 
correction for a series of exposure times. You can see that the noise level of the raw data 
(red) increases significantly with increasing exposure time. With dark spectrum correction 
(green)  however,  the  noise  level  should  remain  pretty  much  at  the  same  level  for