Keysight Technologies U1583BClip-on ammeter adapter 0.5 -400 A 32 mm U1583B Benutzerhandbuch

Produktcode
U1583B
Seite von 4
Current Clamp with
BNC connector
BNC-to-banana 
plug
Handheld Digital 
Multimeter (DMM)
Current Clamp with
BNC connector
Handheld 
Oscilloscope
 
 
  Figure 1: Connection Diagram
Alignment Marks
Put one conductor only within the jaws on inside section of the 
indicated marks as much as possible in order to meet the accuracy of 
specifi cations. Make sure the clamp is vertical to the conductor.
Current Range Selection
To push and release the yellow button for 40 A range and lock the button for 400 A range.
Button
Range
Output
Release  
~ 40 A
~ 10 mV/A
Lock  
~ 400 A
~ 1 mV/A
Operation
AC current can be measured without removing the conductor out of the circuit by following the procedures 
as shown below:
1. 
Plug the BNC of cable to the BNC with dual banana plugs, and then plug into the V/COM terminals on 
 
multimeter. For oscilloscope, plug the BNC connector directly to oscilloscope.
2. 
Set the ACV measurement and range on multimeter you used.
3. 
Position the jaw to a single conductor and center it accordingly to alignment marks.
4. 
Ensure that the arrow marked on the clamp jaw point towards the load of phase measurements or 
 
away from the load (toward the source) for neutral measurements.
5. 
Observe the AC value on a multimeter or the waveform on an oscilloscope which is proportional to the 
 current.
6. 
Select a lower range on the Clamp-on adapter and set the corresponding sensitivity (mV/ A setting) 
 
on the oscilloscope if required.
Calibration Equipment 
        
The pre-calibration guidelines are shown as follows:
• 
Be sure you  are a qualifi ed person to perform the calibration
• 
The environment should be 23 °C ±2 °C and the relative humidity (RH) shall be < 80%.
The test equipment requirements listed in table below or equivalents are required to perform the calibration 
and performance verifi cation test procedures. Alternative equipment may be used as long as the accuracy is 
as good as or better than the specifi cations listed.
Standard Source
Operating Range
Accuracy 
Required
Recommended 
Equipment
AC Current Calibrator
33 mA – 329.99 mA at 10 Hz to 3 kHz
0.33 A – 2.99999 A at 10 Hz to 3 kHz
3 A – 20.5 A at 10 Hz to 3 kHz
≤ ± 0.2 %
≤ ± 0.6 %
≤ ± 3.0 %
Wavetek 9100 or Fluke 5520A or 5101B 
or equivalent
Multimeter
AC 500.0 mV or 1000.0 mV
≤ ± 1.5 %
Agilent U1251A or U1252A or
Agilent-34405A or equivalent
50 Turns Current Coil
0.2 A – 20.5 A
≤ ± 1.0 %
Fluke 5500A Coil or Wavetek 9100 
Option 200 or equivalent