3.  Press Attenuation on the side menu 
and use the Variable knob to set 
the attenuation. 
Alternatively, press Set to 10X.  
 
 
 
Range 
1mX ~1kX (1-2-5 step) 
Note 
The attenuation factor adds no influence on the real 
signal. It just changes the voltage/current scale on the 
display. 
Set the Deskew 
 
Background 
The deskew function is used to compensate for the 
propagation delay between the oscilloscope and 
the probe. 
Panel Operation 
1.  Press one of the Channel keys. 
CH1
 
2.  Press Probe from the bottom menu.  
 
3.  Press Deskew on the side menu and 
use the Variable knob to set the 
deskew time.  
Alternatively, press Set to 0s to 
reset the deskew time. 
 
 
 
Range 
-50ns~50ns, 10ps increments  
 
4.  Repeat the procedure for another channel if 
necessary.