Gw Instek GDS-2304A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2304A Datenbogen

Produktcode
GDS-2304A
Seite von 261
 
 
CONFIGURATION 
145 
 
Delay trigger example (by time) 
 
A
B
C
D
 
A  Ext. trigger input 
 
B  Source  
 
C  Delay time length 
 
D  First triggering point 
Pulse Width 
Triggers when the pulse width of the signal is less 
than, equal, not equal or greater than a specified 
pulse width. 
 
 
 
Pulse width 
Video 
Extracts a sync pulse from a video format signal, 
and triggers on a specific line or field. 
Pulse and Runt 
Triggers on a “runt”. A runt is a pulse that passes a 
specified threshold but fails to pass a second 
threshold.  Both positive and negative runts can be 
detected. 
 
A
B
C
D
 
A  Pulse 
B  Runt 
C  High threshold 
D  Low threshold 
Rise and Fall 
Trigger on rising and or falling edges, below or 
over a specified rate. The threshold can also be 
specified. 
 
B
A
 
A  Thresholds 
B  Rate (time)