Gw Instek GDS-2304A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2304A Datenbogen

Produktcode
GDS-2304A
Seite von 261
 
 
CONFIGURATION 
147 
Trigger Parameter Overview 
 
 
All the following parameters are common for all 
the trigger types unless stated otherwise. 
Trigger Source 
 
CH1 ~ 4 
Channel 1 ~ 4 input signals 
EXT 
External trigger input 
signal 
EXT  TRIG
 
 
AC Line 
AC mains signal 
 
 
Alternate 
Alternate between channel sources for 
the trigger source. 
 
EXT Probe 
Probe trigger source. Set the probe as 
either current or voltage. 
Trigger Mode 
 
Auto (un-
triggered 
roll) 
The GDS-2000A generates an internal 
trigger if there is no trigger event, to 
make sure waveforms are constantly 
updated regardless of trigger events. 
Select this mode especially when 
viewing rolling waveforms at slower 
timebases. 
 
Normal 
The GDS-2000A acquires a waveform 
only when a trigger event occurs. 
 
Single 
The GDS-2000A acquires 
a waveform once when a 
trigger event occurs, then 
stops acquiring. Press the 
Single key to acquire a 
waveform again. 
Single