Gw Instek AFG-2025 Function Generator, Frequency Generator AFG-2025 Datenbogen

Produktcode
AFG-2025
Seite von 145
 
 
AFG-2000 Series User Manual 
86
 
Functionality 
check 
Run this query command via the terminal. 
*idn? 
This should return the Manufacturer, Model 
number, Serial number, and Firmware version in 
the following format. 
GW INSTEK, AFG-2125, SN:XXXXXXXX,Vm.mm 
Note 
^j and ^m can be used as the terminal character 
when using a terminal program. 
PC Software 
 
The proprietary PC software, downloadable from 
GWInstek website, can be used to download 
waveforms.  
Command Syntax 
 
Compatible 
standard 
 
IEEE488.2, 1992 (fully compatible) 
 
SCPI, 1994 (partially compatible) 
Command Tree 
The SCPI standard is an ASCII based standard that 
defines the command syntax and structure for 
programmable instruments.  
 
Commands are based on a hierarchical tree 
structure. Each command keyword is a node on 
the command tree with the first keyword as the 
root node. Each sub node is separated with a 
colon. 
 
Shown below is a section of the SOURce[1]  root 
node and the APPLy/OUTPut and 
SINusoid/SQUare sub nodes.