TEST AFE5809
Figure 7. TSW1400 GUI Configuration
•
From the AFE5809 GUI, go to the ADC page, and then select Ramp.
•
Press the Capture button on the TSW1400 GUI. This displays a ramping waveform on the TSW1400
GUI display area as shown in
GUI display area as shown in
•
Verify that Maximum value is 16383 for 14-bit mode.
•
Repeat for Channel 2 and Channel 8.
•
If each channel has the output as shown in
, proceed with the next step; otherwise, contact the
TI FAE to troubleshoot the problem.
•
On the AFE5809 GUI, change Test Pattern from Ramp to None for the next step.
9
SLOU338A – October 2012 – Revised November 2012
AFE5809EVM Evaluation Module