Running the Software
Figure 13. Device Manager Screen
The USB driver installation is now complete and the EVM is ready to use.
4
Running the Software
Run the GUI software from the Start menu by selecting
All Programs
Texas Instruments
AFE44x0SPO2EVM GUI. Unless the hardware has been
disconnected, observe messages that confirm the connection has been established and the program waits
in idle mode for user input.
If the connection to the AFE44x0SPO2EVM board is not established, the program prompts to continue to
run the GUI in Simulation mode, or to Stop and Close the GUI and check if the AFE44x0SPO2EVM is
connected to the PC.
Figure 14. AFE44x0SPO2EVM Not Connected Error Message
4.1
Overview of the Features
This section provides a quick overview of the various features and functions of the AFE44x0SPO2EVM
software GUI. The GUI allows the user to easily configure the various functions of the AFE such as the
receiver gain and bandwidth settings and transmitter or LED current settings, and timing and clocking
control settings. The GUI supports both AFE4400 and AFE4490 devices. Features not available for the
AFE4400 device are disabled and are not shown in the AFE4400SPO2EVM GUI.
Operations in the GUI should only be performed after the status bar (located at the bottom of the GUI)
displays Ready For New Command.
The main tabs consist of:
About – Product Safety Warnings, Restrictions and Disclaimers (see
Device Configuration – Configures all the AFE44x0 user registers in a series of related subtabs.
Global Settings
Tx Stage
Rx Stage
Timing Controls
12
AFE4400 and AFE4490 Development Guide
SLAU480A – January 2013 – Revised February 2013