Texas Instruments AFE4400 Evaluation Module AFE4400SPO2EVM AFE4400SPO2EVM Datenbogen

Produktcode
AFE4400SPO2EVM
Seite von 59
Running the Software
Low Level Configuration
ADC Capture & Analysis – For viewing and analyzing the raw data.
% SpO2 & HR Display – Provides streaming of oxygen saturation signal along with % SpO2 and heart
rate.
Save – For writing data samples and analysis results to a file.
Figure 15. Product Safety Warnings, Restrictions and Disclaimers
4.1.1
Device Configuration Tab
The Device Configuration tab allows configuration of the various registers of the AFE44x0 device. This
subtab contains five subtabs: Global SettingsTx StageRx StageTiming Controls and Low Level
Configuration
.
4.1.1.1
Global Settings Subtab
The Global Settings subtab for the AFE4490 device shown in
and for AFE4400 device shown in
has the following features:
1. View the Device ID and Firmware Revision
2. Device Reset button that resets the device. (Please note that after a device reset is issued, the
AFE44x0 device registers must be programmed correctly for the PC application GUI to function
properly. See Reset to EVM Defaults on how to issue a device reset and also program the AFE44x0
registers to the EVM default register settings)
3. Reset to EVM Defaults button that resets the device and sets up the board to the EVM default
register settings.
4. Enables the user to set or reset:
(a) SPI Read
(b) XTAL Disable
(c) En Bypass ADC (available for AFE4490 device only)
13
SLAU480A – January 2013 – Revised February 2013
AFE4400 and AFE4490 Development Guide
Copyright © 2013, Texas Instruments Incorporated