Keithley 2182A/E Digital-Multimeter, DMM, 2182A/E Datenbogen

Produktcode
2182A/E
Seite von 6
LO
W L
E
V
E
L M
E
A
S
U
R
E & S
O
U
R
C
E
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY 
(U.S. only)
M
ode
l 2
18
2A s
pe
ci
fic
at
io
ns
Volts specifications 
(20% over range)
CONDITIONS: 1PLC with 10 reading digital filter or 5PLC with 2 reading digital filter .
accuracy: ±(ppm of reading + ppm of range) 
(ppm = parts per million) (e.g., 10ppm = 0.001%)
Temperature
Channel 1 
Input 
24 Hour
 1
 
90 Day 
1 Year 
2 Year 
Coefficient 
 range 
resolution 
resistance 
T
Cal
 ±1°C 
T
Cal
 ±5°C 
T
Cal
 ±5°C 
T
Cal
 ±5°C 
0°–18°C & 28°–50°C
10 .000000 mV
 2, 3, 4
1 nV 
>10  G
20 + 4 
40 + 4 
50 + 4 
60 + 4 
(1 + 0 .5)/°C
100 .00000  mV 
10 nV 
>10  G
10 + 3 
25 + 3 
30 + 4 
40 + 5 
(1 + 0 .2)/°C
1 .0000000 
100 nV 
>10  G
7 + 2 
18 + 2 
25 + 2 
32 + 3 
(1 + 0 .1)/°C
10 .000000 
1 µV 
>10  G
2 + 1
 5
 
18 + 2 
25 + 2 
32 + 3 
(1 + 0 .1)/°C
  100 .00000 V
 4
10 µV 
10  M
W ±1% 
10 + 3 
25 + 3 
35 + 4 
52 + 5 
(1 + 0 .5)/°C
Channel 2
 6, 10
100 .00000  mV 
10 nV 
>10  G
10 + 6 
25 + 6 
30 + 7 
40 + 7 
(1 + 1   )/°C
1 .0000000 
100 nV 
>10  G
7 + 2 
18 + 2 
25 + 2 
32 + 3 
(1 + 0 .5)/°C
10 .000000 
1 µV 
>10  G
2 + 1
 5
 
18 + 2 
25 + 2 
32 + 3 
(1 + 0 .5)/°C
CHANNEL 1/CHANNEL 2 RATIO: For input signals 
≥1% of the range, Ratio Accuracy = 
±{[Channel 1 ppm of Reading + Channel 1 ppm of Range * (Channel 1 Range/Channel 1 Input)] + [Channel 2 ppm of Reading + Channel 2 ppm of Range * (Channel 2 Range/Channel 2 Input)]} .
DELTA (hardware-triggered coordination with Series 24XX, Series 26XXA, or Series 622X current sources for low noise R measurement): 
Accuracy = accuracy of selected Channel 1 range plus accuracy of I source range .
DELTA MEASUREMENT NOISE WITH 6220 or 6221: Typical 3nVrms / Hz  (10mV range)
21
 . 1Hz achieved with 1PLC, delay = 1ms, RPT filter = 23 (20 if 50Hz) .
PULSE-MODE (WITH 6221): Line synchronized voltage measurements within current pulses from 50µs to 12ms, pulse repetition rate up to 12Hz .
PULSE MEASUREMENT NOISE (typical rms noise, R
DUT
<10
W): ±(0 .009ppm of range*) / meas_time / pulse_avg_count + 3nV** /   (2 · meas_time · pulse_avg_count)  for 10mV range .
* 0 .0028ppm for the 100mV range, 0 .0016ppm for ranges 1V and above .
** 8nV/ Hz for ranges above 10mV . meas_time (seconds) = pulsewidth – pulse_meas_delay in 33µs incr .
DC Noise Performance
 7
 (DC noise expressed in volts peak-to-peak)
Response time = time required for reading to be settled within noise levels from a stepped input, 60Hz operation .
Channel 1
 response 
range 
Time 
NPlC, filter 
10 mV 
100 mV 
1 V 
10 V 
100 V 
NMrr
 8
CMrr
 9
25 .0 
5, 75 
6 nV 
20 nV 
75 nV 
750 nV 
75 µV 
110 dB 
140 dB
4 .0 
5, 10 
15 nV 
50 nV 
150 nV 
1 .5 µV 
75 µV 
100 dB 
140 dB
1 .0 
1, 18 
25 nV 
175 nV 
600 nV 
2 .5 µV 
100 µV 
95 dB 
140 dB
667  ms 
1, 10 or 5, 2 
35 nV 
250 nV 
650 nV 
3 .3 µV 
150 µV 
90 dB 
140 dB
60  ms 
1, Off 
70 nV 
300 nV 
700 nV 
6 .6 µV 
300 µV 
60 dB 
140 dB
Channel 2
 6, 10
25 .0 
5, 75 
— 
150 nV 
200 nV 
750 nV 
— 
110 dB 
140 dB
4 .0 
5, 10 
— 
150 nV 
200 nV 
1 .5 µV 
— 
100 dB 
140 dB
1 .0 
1, 10 or 5, 2 
— 
175 nV 
400 nV 
2 .5 µV 
— 
90 dB 
140 dB
85  ms 
1, Off 
— 
425 nV 
1 µV 
9 .5 µV 
— 
60 dB 
140 dB
VolTaGE NoIsE Vs. sourCE rEsIsTaNCE
 11
(DC noise expressed in volts peak-to-peak)
 source
analog
Digital 
 resistance
Noise 
filter 
filter
6 nV 
Off 
100
100 
8 nV 
Off 
100
1 k
15 nV 
Off 
100
10 k
35 nV 
Off 
100
100 k
100 nV 
On 
100
1 M
350 nV 
On 
100
TEMPEraTurE (Thermocouples)
 12
aCCuraCY 
(Displayed in °C, °f, or K. accuracy based on 
90 Day/1 Year 
 ITs-90, exclusive of thermocouple errors.)  
23° ±5°C 
 
relative to simulated 
 TYPE 
raNGE 
rEsoluTIoN reference Junction
–200 to  +760°C
0 .001 °C 
±0 .2 °C
–200 to  +1372°C
0 .001 °C 
±0 .2 °C
–200 to +1300°C
0 .001 °C 
±0 .2 °C
–200 to  +400°C
0 .001 °C 
±0 .2 °C
–200 to +1000°C
0 .001 °C 
±0 .2 °C
0 to  +1768°C 
0 .1 °C 
±0 .2 °C
0 to  +1768°C 
0 .1 °C 
±0 .2 °C
+350 to  +1820°C 
0 .1 °C 
±0 .2 °C
2182a
Nanovoltmeter
operating Characteristics
 13, 14
 
60Hz (50Hz) operation
 function 
Digits 
readings/s 
PlCs
DCV Channel 1, 
7 .5 
3 (2) 
5
Channel 2,
7 .5
 17, 19
6 (4) 
5
Thermocouple
6 .5
 18, 19
18 (15) 
1
6 .5
 18, 19, 20
45 (36) 
1
5 .5
 17, 19
80 (72) 
0 .1
4 .5
 16, 17, 19
115 (105) 
0 .01
Channel 1/Channel 2 (Ratio), 
7 .5 
1 .5 (1 .3) 
5
Delta with 24XX, Scan
7 .5
 17, 19
2 .3 (2 .1) 
5
6 .5
 18
8 .5 (7 .5) 
1
6 .5
 18, 20
20 
 (16) 
1
5 .5
 17
30   (29) 
0 .1
4 .5
 17
41 (40) 
0 .01
Delta with 622X 
6 .5 
47 (40 .0)
 22
1
system speeds
 13, 15
RANGE CHANGE TIME:
 14
 
<40 ms  (<50 ms) .
FUNCTION CHANGE TIME:
 14
 
<45 ms  (<55 ms) .
AUTORANGE TIME:
 14
 
<60 ms  (<70 ms) .
ASCII READING TO RS-232 (19.2K Baud): 40/s  (40/s) .
MAX. INTERNAL TRIGGER RATE:
 16
 
120/s (120/s) .
MAX. EXTERNAL TRIGGER RATE:
 16
 
120/s (120/s) .