4-22
Range, Digits, Rate, Bandwidth, and Filter
Filter programming examples
Example #1 — The following command sequence configures filtering for the DCI 
function:
CURR:TCON MOV
' Select the moving filter.
CURR:AVER:WIND 0.01
' Set filter window to 0.01%.
CURR:AVER:COUN 10
' Set to filter 10 readings.
CURR:AVER ON
' Enable filter.
Example #2 — The following command sequence configures channels 101 through 115 
of the Model 7700 to use the repeat filter when they are scanned.
FUNC 'VOLT'
' Select DCV function.
VOLT:AVER:TCON REP
' Select the repeating filter.
VOLT:AVER:COUN 20,(@101:115)
' Set to filter 20 readings.
VOLT:AVER ON,(@101:115)
' Enable filter.