1-13
Rack mounting
The Model 2750 can be mounted in a standard 19-inch rack. For a mainframe that has one 
or no switching modules installed, the Model 2750 can be secured to the front rails of the 
rack.
The hardware necessary to secure the mainframe to the front rails of the rack is supplied 
with the Model 2750. The supplied hardware kit includes four dress screws and four 
retaining clips. Each retaining clip has a captive nut.
Perform the following steps to mount the Model 2750 to front rails of the rack:
WARNING
Make sure the Model 2750 is turned off, the line cord is disconnected, 
and it is not connected to any external circuitry.
1.
Select a location in the rack. The mainframe takes up 3H-inches of vertical space.
2.
Using 
 as a guide, install two retaining clips on the left front rail. Slide 
each retaining clip over a mounting hole such that the captive nut is positioned on 
the inside of the rack cabinet. In a similar manner, install two retaining clips on the 
right front rail.
3.
Remove the four foot assemblies from the bottom of the Model 2750. The retaining 
screw for an assembly is located under the rubber foot. Simply pull off the rubber 
feet to gain access to the screws. Retain these foot assemblies for future use.
4.
Position the Model 2750 in the rack and loosely attach the front panel to the rack 
rails using the four supplied dress screws.
5.
Tighten the four dress screws.
CAUTION
For a Model 2750 that has two or more installed switching modules, 
rear brackets are required to support the additional weight. The Model 
4288-7 is a rack mount kit for the Model 2750 that uses rear support 
brackets.
Figure 1-3
Rack preparation
1.75”
Left Front
Rack Rail
Retaining Clips
with Nuts
2750-900-01.book  Page 13  Wednesday, August 3, 2011  7:56 AM