PEL-3000 Programming Manual 
52
 
Query example 
:GNG? 
>GO 
Returns the Go-NoGo test result. 
 [:CONFigure]:GNG:H 
Set
Query
 
 
Description 
Sets or queries the high voltage/current limit 
value. If the entry mode is set to value, the high 
voltage/current limit value units are in 
volts/amps. If the entry mode is set to percent, the 
high voltage/current limit value units are in 
percent.  
Syntax 
[:CONFigure]:GNG:H <NRf> 
Query Syntax 
[:CONFigure]:GNG:H? 
Parameter 
<NRf> 
Sets the high 
voltage/current limit value 
in volts/amps or in 
percent. 
Return parameter  <NR2> 
Returns the high 
voltage/current limit value 
in volts/amps or as %. 
Example 
:GNG:H 100.0 
Sets the high voltage limit value to 100%. 
Query example 
:GNG:H? 
>100.0 
Returns the high voltage limit value as 100.0%. 
[:CONFigure]:GNG:L 
Set
Query
 
 
Description 
Sets or queries the low voltage/current limit value. 
If the entry mode is set to value, the low 
voltage/current limit value units are in 
volts/amps. If the entry mode is set to percent, the 
low voltage/current limit value units are in 
percent.  
Syntax 
[:CONFigure]:GNG:L <NRf>