PEL-3000 Programming Manual 
54
 
Parallel Commands 
 
 
[:CONFigure]:PARallel ........................................................ 54
 
 
[:CONFigure]:PARallel 
Set
Query
 
 
Description 
Configures the unit for parallel operation, or 
queries its state. This command configures the unit 
as a Master or Slave, and configures how many 
slave units are connected if the unit is configured 
as a master. 
Note 
The booster (PEL-3211) cannot use low current 
range. 
Syntax 
[:CONFigure]:PARallel { OFF | MASTer | SLAVe 
|P2|P3|P4|P5|B1|B2|B3|B4 } 
Query Syntax 
[:CONFigure]:PARallel? 
Parameter 
P2,P3,P4 or P5 
 
B1,B2,B3 or B4 
Number of connected 
slaves 
Number of connected 
booster 
 
OFF 
Turn parallel mode off 
 
MASTER or MAST 
Sets the unit to Master 
SLAVE or SLAV 
Sets the unit to Slave 
Return parameter  <ASCII string> 
Returns an ASCII string 
with the mode of the unit 
(Master/Slave) and the 
number of connected 
devices. 
 
Example 
:PARallel MAST 
Sets the unit to Master. 
:PARallel B2 
Configures the unit for use with 2 booster units.