Gw Instek PEL-3211 Electronic Load PEL-3211 Datenbogen
Produktcode
PEL-3211
COMMNAND OVERVIEW
69
Parameter
FAST
Fast response
SLOW
Slow response
Return parameter <ASCII string>
Returns CV response
speed as a string.
speed as a string.
Example
:RESPonse FAST
Sets the CV response to fast.
Sets the CV response to fast.
Query example
:RESPonse?
> CV mode response:FAST.
> CV mode response:FAST.
[:MODE]:DYNamic
Set
Query
Description
Sets or queries the switching mode. Dynamic is the
default mode.
default mode.
Syntax
:MODE:DYNamic{ DYNamic | STATic }
Query Syntax
:MODE:DYNamic?
Parameter
DYNAMIC or DYN
Set to Dynamic mode
STATIC or STAT
Set to Static mode
Return parameter <ASCII string>
Returns the switching
mode.
mode.
Example
:MODE:DYNamic DYN?
Set the switching mode to dynamic
Set the switching mode to dynamic
Query example
:MODE:DYNamic?
> Dynamic, ;Dynamic Level:Value, Dynamic
Time:T1/T2
The switching mode is set to dynamic mode.
> Dynamic, ;Dynamic Level:Value, Dynamic
Time:T1/T2
The switching mode is set to dynamic mode.