OPERATION 
63 
Basic Configuration 
The basic configuration settings are the common configuration 
settings that are used for each operating mode. After selecting a 
basic operating mode (CC, CR, CV or CP mode), the slew rate, 
switching mode, response rate and other common parameters 
should be configured.  
Select the Switching Function  
 
Description 
The PEL-3000 has two switching modes, static 
and dynamic. The switching modes allow the 
PEL-3000 to switch between two preset levels. 
Static mode can only switch between the two 
levels manually, while Dynamic mode switches 
between each level automatically based on a 
timer. 
 
Static mode: A Value, B Value 
 
Dynamic mode: Level1, Level2  
When the unit is set to static mode, only one 
value (A Value or B Value) can be active at a 
time. The active value is shown in the 
Operation Status Panel. 
 
 
Function
Static
Configure
0.00
w
LOAD
RS232
SHORT
A Value
Fine
W
W
Active setting
 
When the unit is set to dynamic mode, the unit 
will switch between Level1 and Level2 based 
on the Timer1 and Timer2 parameters, shown 
below.