Gw Instek PEL-3111 Electronic Load PEL-3111 Datenbogen

Produktcode
PEL-3111
Seite von 119
 
 
PEL-3000 User Manual 
64
 
Level
Time
Level 2
Level 1
Timer1 Timer2 Timer1
Dynamic Mode
 
 Note 
Dynamic mode is not available for CV mode. 
Operation 
1.  Make sure the load is off. 
2.  Press 
Main
.  
3.  Select Dynamic or Static mode with the 
Function[F4] soft-key.  
 
A different switching mode can be set for CC, 
CR and CP mode. 
 
4.  For dynamic mode, set the Timer1 and Timer2 
parameters using the scroll wheel and number 
pad.  
 
Timer1 sets the Level1 on-time. 
 
Timer2 sets the Level2 on-time. 
 
Take the slew rate settings into consideration 
when setting the timers. 
 
The frequency of the dynamic switching is 
output via the TRIG OUT BNC.  
 
To select whether A Value or B Value is the 
―active‖ setting, press the 
Shift
+
Preset
 
keys. 
 
The ―active‖ value will be shown in the 