OPERATION 
89 
Go-NoGo  
The Go-NoGo configuration is used to create pass/fail limits on the 
voltage or current input. If the voltage/current exceeds the 
pass/fail limits, an alarm will be output.  
The Go-NoGo configuration can be used with the Program function 
to create complex pass/fail tests.  
Setting the Go-NoGo Limits 
 
Description 
The Go-NoGo setting limits can be set as either 
discrete high & low values or as a percentage 
offset from a center value. 
Operation 
1.  Press 
Main
 > Configure[F5] > Go-NoGo[F3]. 
2.  Select Entry Mode and choose how to set the 
pass/fail limits. 
 
Value will allow you to set the limits as discrete 
values. 
 
Percent will allow you to set the limits as a 
percentage offset from a center value. 
 
3.  If Entry Mode was set to Value, Set the High & 
Low limit values.  
 
High: 
Low: 
0~rating current/voltage 
0 ~rating current/voltage 
 
4.  If Entry Mode was set to Percent, Set the Center 
voltage/current and High, Low % values.  
 
Center: 
High: 
Low: 
0~rating current/voltage 
0~100% of center voltage/current 
0~100% of center voltage/current