Go-NoGo
The Go-NoGo configuration is used to create pass/fail limits on the
voltage or current input. If the voltage/current exceeds the
pass/fail limits, an alarm will be output.
voltage or current input. If the voltage/current exceeds the
pass/fail limits, an alarm will be output.
The Go-NoGo configuration can be used with the Program function
to create complex pass/fail tests.
to create complex pass/fail tests.
Setting the Go-NoGo Limits
Description
The Go-NoGo setting limits can be set as either
discrete high & low values or as a percentage
offset from a center value.
discrete high & low values or as a percentage
offset from a center value.
Operation
1. Press
Main
> Configure[F5] > Go-NoGo[F3].
2. Select Entry Mode and choose how to set the
pass/fail limits.
Value will allow you to set the limits as discrete
values.
values.
Percent will allow you to set the limits as a
percentage offset from a center value.
percentage offset from a center value.
3. If Entry Mode was set to Value, Set the High &
Low limit values.
High:
Low:
Low:
0~rating current/voltage
0 ~rating current/voltage
0 ~rating current/voltage
4. If Entry Mode was set to Percent, Set the Center
voltage/current and High, Low % values.
Center:
High:
Low:
High:
Low:
0~rating current/voltage
0~100% of center voltage/current
0~100% of center voltage/current
0~100% of center voltage/current
0~100% of center voltage/current