Load Line 
Inductance 
Considerations  
When using the PEL-3000 load generator, 
voltage drop and voltage generated due to load 
line inductance and current change must be 
taken into account. Extreme changes in voltage 
may exceed the minimum or maximum voltage 
limits. Exceeding the maximum voltage limit 
may damage the PEL-3000.  
 
 To determine the voltage generated, the 
following equation can be used.  
E = L x (∆ I / ∆ T) 
E= voltage generated 
L=load line inductance 
∆ I= change of current (A) 
∆ T= time (us)   
Load line inductance (L) can be approximated 
as 1uH per 1 meter of wire. (∆ I / ∆ T) is the 
slew rate in A/us.