Guidance and manufacturer’s declaration – electromagnetic immunity
The “Mini Daylight” is intended for use in the electromagnetic environment specified below. The customer or
the user of the “Mini Daylight” should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 V
3 V/m
Portable and mobile RF communications 
equipment should be used no closer to any 
part of the “Mini Daylight”, including cables, 
than the recommended separation distance 
calculated from the equation applicable to 
the frequency of the transmitter.
Recommended separation distance
d=1.2   P
d=1.2   P   80MHz to 800MHz
d=2.3   P  800MHz to 2.5 GHz
where P is the maximum output power rating 
of the transmitter in watts (W) according 
to the transmitter manufacturer and d is 
the recommended separation distance in 
metres (m).
Field strengths from fixed RF transmitters, 
as determined by an electromagnetic site 
survey, 
should be less than the compliance 
level in each frequency range.
b
Interference may occur in the vicinity of 
equipment marked with the following 
symbol:
 
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorpti-
on and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic 
site survey should be considered. If the measured field strength in the location in which the “Mini Daylight” is 
used exceeds the applicable RF compliance level above, the Medical Daylight should be observed to verify 
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reori-
enting or relocating the “Mini Daylight”.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.