Guidance & Declaration — electromagnetic immunity
The model TL80 is intended for use in the electromagnetic environment specified below. The customer 
or the user should assure that It is used in such an environment.
immunity test
iEc 60601
test level
compliance level
Electromagnetic environment 
- guidance
Electrostatic 
discharge (ESD) 
lEC 61000-4-2 
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors 
are covered with synthetic 
material, the relative humidity 
should be at least 30 %.
Electrical fast tran-
sient/burst 
IEC 61000-4-4
±2kV for power sup-
ply lines
±1 kV for Input/out-
put lines
±2kV for power supply 
lines
Mains power quality should be 
that of a typical commercial or 
hospital environment.
Surge 
lEC 61000-4-5 
±1 kV line to line
±2 kV line to earth
±2 kV line to earth
Mains power quality should be 
that of a typical commercial or 
hospital environment.
Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines 
IEC 61000-4-11.
<5 % U
T
(>95% dip in U
T
.)
for 0.5 cycle 
40 % U
T
 
(60% dip in U
T
for 5 cycles 
70% U
T
 
(30% dip in U
T
for 25 cycles 
<5% U
T
(>95 % dip in U
T
for 5 sec 
<5 % U
T
(>95% dip in U
T
.)
for 0.5 cycle 
40 % U
T
 
(60% dip in U
T
for 5 cycles 
70% U
T
 
(30% dip in U
T
for 25 cycles 
<5% U
T
(>95 % dip in U
T
for 5 sec
Mains power quality should be 
that of a typical commercial 
or hospital environment. If the 
users of the model TL80 require 
continued operation during 
power mains interruptions, it is 
recommended that the model 
TL80 be powered from an un-
interruptible power supply or a 
battery.
Power frequency 
(50/60 Hz)
magnetic field 
lEC 61000-4-8 
3 A/m
Not applicable
Not applicable
NOTE  
U
T
  is the a.c. mains voltage prior to application of the test level.