recommended separation distances between 
portable and mobile rf communications equipment and the model Tl80
The models TL80 are intended for use in electromagnetic environment in which radiated RF distur-
bances is controlled. The customer or the user of the model TL80 can help prevent electromagnetic 
interference by maintaining a minimum distance between portable and mobile RF communications 
equipment (transmitters) and the model TL80 as recommended below, according to the maximum out-
put power of the communications equipment.
rated maximum out-
put power 
of transmitter 
W
Separation distance according to frequency of transmitter
m
150kHz to 80mHz
d=1.2×P
1/2
80mHz to 800mHz
d=1.2×P
1/2
800mHz to 2,5GHz
d=2.3×P
1/2
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation 
distance d in meters (m) can be estimated using the equation applicable to the frequency of the trans-
mitter, where P is the maximum output power rating of the transmitter in watts (W) accordable to the 
transmitter manufacturer.
NoTE i At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies. 
NoTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people.