Beurer TL30 608.05 Datenbogen

Produktcode
608.05
Seite von 13
66
Table 4: Recommended separation distances between portable and mobile RF communications 
equipment and the model TL30
Recommended separation distances between 
portable and mobile RF communications equipment and the model TL30
The model TL30 is intended for use in electromagnetic environment in which radiated RF disturbances is 
controlled. The customer or the user of the model TL30 can help prevent electromagnetic interference by 
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) 
and the model TL30 as recommended below, according to the maximum output power of the communications 
equipment.
Rated maximum output 
power of transmitter 
W
Separation distance according to frequency of transmitter 
m
150 kHz to 80 MHz
1.2   P
80 MHz to 800 MHz
1.2   P
800 MHz to 2,5 GHz
2.3   P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d 
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is 
the maximum output power rating of the transmitter in watts (W) accordable to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies. 
NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorpti-
on and reflection from structures, objects and people.