Technical Description Concerning Electromagnetic Emission
Table 1: Declaration - electromagnetic emissions
Guidance and manufacturer’s declaration – electromagnetic immunity
The model TL30 is intended for use in the electromagnetic environment specified below.  
The customer or the user should assure that It is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic 
discharge (ESD) 
lEC 61000-4-2 
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete 
or ceramic tile. If floors are covered 
with synthetic material, the relative 
humidity should be at least 30 %. 
Electrical fast 
transient/burst 
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output 
lines
±2 kV for power
supply lines
Mains power quality should be that 
of a typical commercial or hospital 
environment.
Surge 
lEC 61000-4-5 
±1 kV line to line
±2 kV line to earth
±2 kV line to earth
Mains power quality should be that 
of a typical commercial or hospital 
environment.
Voltage dips, 
short interrup-
tions and vol-
tage variations 
on power supp-
ly input lines 
IEC 61000-4-11
<5 % U
T
 (>95% dip in U
T
)
for 0.5 cycle 
40 % U
T
 (60% dip in U
T
for 5 cycles 
70% U
T
 (30% dip in U
T
for 25 cycles 
<5% U
T
 (>95 % dip in U
T
for 5 sec 
<5 % U
T
 (>95% dip in U
T
)
for 0.5 cycle 
40 % U
T
 (60% dip in U
T
for 5 cycles 
70% U
T
 (30% dip in U
T
for 25 cycles 
<5% U
T
 (>95 % dip in U
T
for 5 sec
Mains power quality should be that 
of a typical commercial or hospital 
environment. If the user of the 
model TL30 require continued ope-
ration during power mains interrup-
tions, it is recommended that the 
model 168 DAYLUX Therapy Light 
be powered from an uninterruptible 
power supply or a battery.
Power frequen-
cy (50/60 Hz)
magnetic field 
lEC 61000-4-8
3 A/m
Not applicable
Not applicable
NOTE U
T
 is the a.c. mains voltage prior to application of the test level.