Chapter 3. Programming Site Data
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
65
3.11.1 Entering Correction Factors
The final option in the Program Menu, Correction Factors, allows you to 
enter and modify three correction factors: Reynolds Correction, Kinematic 
Viscosity and Calibration Factor. To enter the Correction Factors option, 
scroll to the Correction Factors entry on the Program Menu and press 
[ENTER]
. The screen appears similar to Figure 35 below. To step through 
each parameter, press the [T] key. Refer to Figure 152 on page 236 o
Appendix A, Menu Maps.
Figure 35: Reynolds Correction Window
3.11.2 Entering Reynolds Correction
The default for Reynolds Correction is “On.” This correction factor should 
be on in most applications, including all those that utilize clamp-on 
transducers. It makes a small adjustment to the flow rate reported by the 
PT878, based on the Kinematic Viscosity. Reynolds Correction is necessary, 
as the velocity of the fluid measured along a diametrical path must be 
related to the total area average velocity over the entire pipe cross-section.
Reynolds Correctio
Inputs
Correction Factors
Cancel
OK
Calibration Factor
Re Correction
On
Off
K/V (v)
m
3
/s
Data Source
K/V Factor   
Single
Table
Edit Table
10