Chapter 3. Programming Site Data
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
67
3.11.3 Entering a Calibration Factor
The Calibration Factor is used to calibrate or adjust the readings of the 
PT878 to another flow reference. To enter a Calibration Factor, return to the 
Reynolds Correction tab at the top of the Correction Factors option. Press 
the [X] arrow key to reach the Calibration Factor window. The window 
appears similar to Figure 36 below. To step through each parameter, press 
the [T] key.
Figure 36: The Calibration Factor Window
1.
The first prompt asks if you wish to enable the calibration factor. 
a.
Use the [
W
] and [
X
] keys to scroll to the appropriate radio button.
b.
Press 
[ENTER]
 to confirm your selection.
2.
The next prompt asks if you want a single K factor or a table of K 
factors.
a.
Use the [
W
] and [
X
] keys to scroll to the appropriate radio button.
b.
Press 
[ENTER]
 to confirm your selection.
Reynolds Correctio
Inputs
Correction Factors
Calibration Factor
Edit Table
Calibration Factor
On
Off
Meter K-Factor
Data Source
K Factor   
Single
Table
10
Edit Table
Cancel
OK