Barnet Ensign Ross ED810 Benutzerhandbuch

Seite von 10
ED810 INSTRUCTION MANUAL  27-02-07 
 
 
MARKING 
 
All units have a rating label which carries the following important information:- 
 
Model type: 
 
GAS DETECTOR TYPE ED810 
 
Input voltage:   
12Vdc/24Vdc   
(delete as appropriate) 
 
Vmax in; 
 
Um: 250V 
 
Code:   
 
[EEx ia] II C 
 
 
Certificate no.   
Baseefa03ATEX0507 
 
Epsilon x, 
 
 
gas group 
 
          II (1) G 
and category: 
 
CE marking, 
Notified body 
number: 
 
 
 
Serial no.  
 
 
 
 
(serial number must be entered here) 
 
Year of construction: 
 
 
 
(year of manufacture must be entered here) 
 
 
 
 
 
GENERAL
 
 
The ED810 is a fixed installation gas detection system which can detect a wide range of 
flammable and toxic gases dependent upon sensor type used.   
 
The ED810 is designed for use in 12V or 24V DC insulated and earthed return electrical 
systems.   For insulated return systems special consideration should be given to the 
“Hazardous Area common connection” terminal.  This should normally be connected to a 
good quality earth suitable for intrinsically safe equipment.  However this terminal is 
internally connected to 0V (-VE) of the DC supply and on insulated return installations may 
produce an earth fault.  Isolation can be provided by using a DC-DC inverter, or if mains is 
available, a suitable linear power supply unit.  Both are available from EDL. 
 
Attention is drawn to the need for correct cabling, particularly to the sensors located in the 
Hazardous Area.  The cable supplied by EDL for use with the ED810 has been tested to 
ensure correct parameters, see installation section of this manual.   
 
Attention is also drawn to the need for calibration soon after installation.  When supplied 
alarm levels are approximately set for the target gas (if stated with order) so that a measure of 
immediate protection is obtained once correct operation has been ensured.  
  
  
 
 
 
 
 
 1180