Eyepiece version
2.1
Converting the microscope
1
Eyepieces with 3D shutter
2
3D-CONTROL DSA-3
3
Line connection
4
3D illumination shutter
Fig. 2
Connections for eyepiece version
1
Fuse compartment
2
Voltage selector
3
Holding unit
4
Window
Fig. 3
Changing the voltage selector on rear of
3D-CONTROL DSA-3
3D-CONTROL DSA-3
Set up the microscope in accordance with
the operating instructions of the Axiovert
100/135/135 M
100/135/135 M
Equip the microscope with LD condenser
0.55 Ph1, Ph2, Ph3, DIC, DIC
Connect the components for 3D microscopy
to the 3D-CONTROL DSA-3 control unit: Fig.
2
Connect eyepieces with 3D shutter to the
2
Connect eyepieces with 3D shutter to the
EYEPIECE sockets
Connect 3D illumination shutter to the
ILLUMINATION SHUTTER socket
Connection to the line:
Prior to connection to the line,
check wether the voltage set on the
selector at the rear of the 3D-
CONTROL DSA-3 unit complies with
the line voltage; it may be necessary to
change from 230 V to 115 V or vice
versa:
selector at the rear of the 3D-
CONTROL DSA-3 unit complies with
the line voltage; it may be necessary to
change from 230 V to 115 V or vice
versa:
Fig. 3
Pull the power plug before changing the
fuses or the voltage
Use a small screwdriver to press out the
clamping piece of the fuse holder at the rear
of the 3D-CONTROL DSA-3 unit; lift the fuse
holder out of the fuse compartment
of the 3D-CONTROL DSA-3 unit; lift the fuse
holder out of the fuse compartment
Remove the fuses
Pull voltage selector out of its holder
Turn voltage selector around 180° and insert
Pull voltage selector out of its holder
Turn voltage selector around 180° and insert
it back in its holder:
the new voltage is visible in the window
the new voltage is visible in the window
Insert the fuses listed under Technical Data
Insert fuse holder back in the fuse
Insert fuse holder back in the fuse
compartment until click-stop