8.5.8  Troubleshooting External Alarm Function 
Procedure 
Table 8-12 explains what to do if the external alarm function does not operate properly.   
 
Table 8-12  Troubleshooting External Alarm Function Failure Symptoms 
Step 
What to do 
How to do it or  
where to find the instructions 
Be sure all the alarm configuration data is 
correct. 
Refer to Section 4 – Configuration, Startup, and 
Operation of Recorder with Display.
 
Check the field wiring. 
Refer to Section 2 – Installation.  
Check that the applicable alarm relay 
actuates properly depending on ALARM 
TYPE configuration selection. 
Alarms can be triggered either by: 
•  reconfiguring the value of the trip points 
(AnSnVA), [n = 1 or 2], or 
•  connecting a signal generator to the input and 
setting the generator level beyond the trip-
point values. 
 
8.5.9  Troubleshooting Remote Switch (Digital Input) Function 
Procedure 
Table 8-13 explains what to do if the digital input function does not operate properly. 
 
Table 8-13  Troubleshooting Remote Switch (Digital Input) Function Failure Symptoms 
Step 
What to do 
How to do it or  
where to find the instructions 
Be sure all remote switch configuration 
data is correct. 
Refer to Section 4 – Configuration, Startup, and 
Operation of Recorder with Display. 
See section for pen parameters set up group or 
remote switch set up group. 
Check the field wiring. 
Refer to Section 2 – Installation.  
Check that the applicable remote switch 
actuates properly depending on REMOTE 
SWITCH configuration selection. 
Remote switch actuates either by: 
•  reconfiguring PENIN for REM1 or REM2, or 
•  reconfiguring REM SW for SP2, MAN, or 
MNFS on REMSW1 or REMSW2.