Temperature Calibration 
The MAX 5000XL temperature 
calibration routines will calibrate and 
verify the accuracy of the oven 
temperature throughout the usable 
range.  The oven is calibrated at twelve 
(12) preset increments and verified at 
temperatures that are user selectable. 
 
 
 
Verification Setup 
The VERIFICATION SETUP menu 
allows the verification of the 
temperature calibration to be enabled 
or disabled, and selection of which 
temperatures are to be verified.  
Selecting ALL CAL POINTS will cause 
verification of twelve (12) preset 
temperatures from 50º C to 575º
 
C.  
Four (4) custom temperatures may be 
entered to verify only those 
temperatures used in your processes.  
Entering a number above 575º C will 
produce a warning message on the 
display and set in the default highest 
temperature of 575º C.  Press [OK] to remove the warning from the display. 
 
Calibration Module Setup 
This menu allows entry of the 
Temperature Calibration Interface 
(TCI) data.  Serial numbers of the 
interface module and probe used for 
the calibration and the certification 
date of the TCI is entered here and 
saved for printing in the various 
reports.