IRIG-B Time Synchronization
~
To use the optional IRIG-B capability you will have to 
choose the type of IRIG-B signal that is connected to 
the IRIG input. This must be selected correctly before 
selecting IRIG-B time synchronizing.
~
Select the required input type (DCLS or 1kHz 
modulated) using the IRIG Standard drop down menu.
~
In the Clock Synchronization section click the IRIG radio 
button.
Update Period
~
The internal clock is synchronized to the currently 
selected reference signal (NTS or IRIG) periodically. 
You can adjust the length of time between updates by 
defining the update period in the entry box provided. 
The default period is 2 hours but the rate may be set 
from a minimum update period of 30 minutes to a 
maximum update period of 24 hours. Having a short 
update period may be required if you are using multiple 
DGy 201 recorders in a time synchronized application.
. . . . . . . . . 