Fluke Lawn Mower 2635A Benutzerhandbuch

Seite von 268
Memory Card Operations
Data File Procedures
3
3-13
Using Data Erase
Perform the procedure in Figure 3-9 to remove a data file from the memory card.
Removing a file does not interrupt the sequential dAtxx file names assigned with the
DATA OPEN command. When dAt99 is reached, the instrument will loop back and
reuse previously assigned file names that have been erased or skipped over. To exit at
any time (file not erased), press the 
C key.
FILES
ENTER
ErASE
OPEn
dAtA
dAt99
dAtxx
dAt00
ErASE
ENTER
ENTER
Selecting the DATA mode.  Press the FILES key 
to access the FILES menu.  Press the up/down 
arrow keys until dAtA is displayed, then press the 
ENTER key.  The menu changes to dAtA.  If an 
error message appears, see Table 3-1.
Selecting the ERASE mode.  Press the 
up/down arrow keys until ERASE is displayed in 
the dAtA menu, then press the ENTER key.  The 
menu changes to ErASE.
Selecting a dAtxx File to ERASE.  Press the 
up/down arrow keys until the dAtxx file to be 
erased is displayed in the ErASE menu, where xx 
represents a number from 00 to 99, then press 
the ENTER key.  The menu changes to SUrE.
Typical
Required
yES
no
SUrE
ENTER
Erasing a dAtxx File.  Press the up/down arrow 
keys to select yES or no in the SUrE menu, then 
press the ENTER key.  yES will erase the file; no 
will cancel the procedure and nothing will be 
erased.  The procedure repeats after pressing 
ENTER, or displays nonE if there are no files to 
erase.  Press the CANCL key to exit.  If an error 
message occurs, see Table 3-1.
FILES
Init
StAt
dir
dAtA
SEtUP
op45f.eps
Figure 3-9. Using DATA ERASE to Delete a Measurement Data File