Configuring a Channel to Measure AC Volts
Perform the procedure in Figure 2-5 to configure a channel for measuring ac volts. In
preparation, the instrument must be in the inactive mode (not scanning or monitoring)
and the desired channel must be selected (see Figure 2-3). To exit at any time (changes
not saved), press the
preparation, the instrument must be in the inactive mode (not scanning or monitoring)
and the desired channel must be selected (see Figure 2-3). To exit at any time (changes
not saved), press the
C key.
Restrictions
Maximum Input. The maximum voltage inputs are 300V ac (rms) for channels 0, 1, 11,
and 150V ac (rms) for channels 2 to 10, and 12 to 20.
and 150V ac (rms) for channels 2 to 10, and 12 to 20.
Frequency. The frequency range for maximum voltage inputs is 20 Hz to 100 Hz. Refer
to Appendix A for derated voltage inputs for frequencies between 100 Hz and 100 kHz.
to Appendix A for derated voltage inputs for frequencies between 100 Hz and 100 kHz.
FUNC
ENTER
SET FUNC
OFF
°
F [
°
C]*
Hz
Ω
VAC
V DC
V DC
Auto
300.00 V **
150.00 V ***
30.000 V
3.0000 V
300.00 mV
300.00 V **
150.00 V ***
30.000 V
3.0000 V
300.00 mV
ENTER
SET FUNC
Typical
Required
Selecting the AC Volts Mode. Press the FUNC
key to access the SET FUNC (Set Function)
menu. Press the up/down arrow keys until VAC
(Volts AC) is displayed, then press the ENTER
key.
key to access the SET FUNC (Set Function)
menu. Press the up/down arrow keys until VAC
(Volts AC) is displayed, then press the ENTER
key.
Selecting the Measurement Scale. Select a
fixed scale or Auto (autoranging). A fixed scale
indicates an upper measurement limit. For
example, the 30.000 V scale measures 30 volts
or less. Measurements beyond the scale limit will
cause an OL (overload) display. The x1 (unity or
x 1) multiplier indicates a reading in volts ac; the
m (milli or x .001) multiplier indicates a reading in
millivolts ac.
fixed scale or Auto (autoranging). A fixed scale
indicates an upper measurement limit. For
example, the 30.000 V scale measures 30 volts
or less. Measurements beyond the scale limit will
cause an OL (overload) display. The x1 (unity or
x 1) multiplier indicates a reading in volts ac; the
m (milli or x .001) multiplier indicates a reading in
millivolts ac.
In Auto (autoranging), the instrument chooses the
scale for the best measurement resolution.
When the scales are changed, the scan is
momentarily slowed.
scale for the best measurement resolution.
When the scales are changed, the scan is
momentarily slowed.
* Depends on
temperature scale.
temperature scale.
** Channels 0, 1,
and 11.
and 11.
*** Channels 2 to
10, and 12 to 20
10, and 12 to 20
.
op17f.eps
Figure 2-5. Configuring a Channel to Measure AC Volts