Front Panel Operations
Configuring a Measurement Channel
2
2-11
Configuring a Channel to Measure Resistance
Perform the procedure in Figure 2-6 to configure a channel for measuring resistance. In
preparation, the instrument must be in the inactive mode (not scanning or monitoring)
and the desired channel must be selected (see Figure 2-3). To exit at any time (changes
not saved), press the 
C key.
The resistance to be measured can be connected using one channel (two-terminal
connection) or two channels (four-terminal connection). The four-terminal connection
provides increased measurement precision. The two channels used in a four-terminal
connection are a decade apart (n and n+10), for example, channel 3 (n) and 13 (n+10).
Only the lower channel is configured.
Restrictions
Four-Terminal Channels. Four-Terminal configurations are limited to channels 1 to 10
(n). The channel a decade higher (n + 10) is automatically reserved for use.
FUNC
ENTER
SET FUNC
OFF
°
F [
°
C]*
Hz
Ω
VAC
V DC
Auto
10.000 M
3.0000 M
300.00 k
30.000 k
3.0000 k
300.00 
ENTER
SET FUNC
Typical
Required
Selecting the Resistance Mode.  Press the 
FUNC key to access the SET FUNC (Set 
Function) menu.  Press the up/down arrow keys 
until 
Ω
 (ohms) is displayed, then press the 
ENTER key.
Selecting the Measurement Scale.  Select a 
fixed scale or Auto (autoranging).  A fixed scale 
indicates an upper measurement limit.  For 
example, the 30.000 k
Ω
 scale measures 30 k
Ω
 
ohms or less.  Measurements beyond the scale 
limit will cause an OL (overload) display.  The x1 
(unity or x 1) multiplier indicates a reading in 
ohms; the k (kilo or x 1,000) multiplier indicates a 
reading in kilohms; the M (mega or x 1,000,000) 
multiplier indicates a reading in megohms.
In Auto (autoranging), the instrument chooses the 
scale for the best measurement resolution.  
When scales are changed, the scan is 
momentarily slowed.
*  Depends on 
temperature scale.
**  Channels 1 to 10.
2T
4T**
SET FUNC
ENTER
Selecting the Terminal Mode.  Select the two-
terminal (2T) or four-terminal (4T) mode.  4T 
automatically clears configuration data from the 
channel a decade higher (n+10).  For example, 
selecting channel 3 for 4T operation will 
automatically clear channel 13 of all configuration 
data and lock it out from further use.
op18f.eps
Figure 2-6. Configuring a Channel to Measure Resistance