Teledyne 401 Benutzerhandbuch

Seite von 114
Teledyne API Model 401 O
3
 Photometric Calibrator Instruction Manual, 01124, Rev. J2
 
Table 7-9:  RS-232 Variable Names 
Var. Name 
Button Sequence 
Legal Values 
AUTO-TIME SETUP-IZSC-TIME 
00:00-23:59 
AUTO-SHIFT 
SETUP-IZSC-SHFT 
-60 - +60 
TIME-ZERO SETUP-IZSC-WAIT 
1-20 
TIME-SPAN SETUP-IZSC-WAIT 
1-20 
O3-SPAN SETUP-IZSC-SPAN  0-9000 
REPORT-FREQ SETUP-DAS-FREQ 
1-60 
MACHINE-ID SETUP-COMM-ID 
0000-9999 
BAUD-RATE 
SETUP-COMM-BAUD 
300, 1200, 2400 
ILAMP-SET (not 
available) 
20-70 
IZSLAMP-SET (not 
available) 
20-70 
LAMP-PCT SETUP-MISC-O3-GEN 0-100 
CURR-TIME SETUP-MISC-CLK-TIME 
00:00-23:59 
CURR-DATE SETUP-MISC-CLK-DATE 
01/01/00-12/31/99 
CLOCK-ADJ 
SETUP-MISC-CLK-ADJ 
-60 - +60 
DA-RANGE SETUP-MISC-D/A-RNGE 
0=100, 1=500, 2=1000, 
3=5000, 4=10000 
DA-OFFSET 
SETUP-MISC-D/A-OFFS 
-500 - +500 
DARK-OFFSET 
SETUP-MISC-DARK-VIEW 
-1000 - +1000 
PASS-ENABLE SETUP-PASS 
OFF, 
ON 
FAULT-TIME SETUP-MISC-MORE-FLT 
0-300 
 
7-10