Teledyne API Model 401 O
3
 Photometric Calibrator Instruction Manual, 01124, Rev. J2 
 
Table 9-1:  Warning Messages 
Warning Message 
Meaning 
Corrective Action 
ANA LAMP 
WARNING 
The O
3
 REF value is greater than 5000 mV 
or less than 2500 mV. 
Check and adjust Source Lamp 
and UV detector as described in 
Section 9.6.6. 
ANA LAMP 
SHUTDOWN 
Temperature control of the Source Lamp 
cannot be maintained at its 52
o
C set point. 
Check source lamp heater and 
thermistor as described in  
Section 8.10. 
O3 GEN LAMP 
WARNING 
The Ozone Generator is unable to produce 
at least 1000 ppb at its maximum output or 
the O
3
 GEN feedback control is adjusting 
the Ozone generator drive signal by more 
than a factor of two. 
Check and adjust the O
GEN 
lamp and reference det as 
described in Section 9.6.8. 
O3 GEN LAMP 
SHUTDOWN 
Temperature control of the O
3
 GEN Lamp 
cannot be maintained at its 48
o
set point. 
Check O
3
 GEN lamp heater and 
thermistor as described in 
Section 8.10. 
SAMPLE 
PRESSURE 
The Sample Pressure is less than 15"Hg or 
is greater than 35"Hg. 
Check for pressure transducer 
problems as described in  
Sections 9.6.1 and 9.6.5. 
SAMPLE FLOW 
WARNING 
The sample flow is less than 500 cc/min or 
greater than 1000 cc/min. Check for 
pneumatic system problems as described 
in Section 9.6.1. 
Check for flow transducer 
problems as described in  
Section 9.6.5. 
BOX TEMP  
WARNING 
The inside chassis temp is less than 5
o
C or 
is greater than 50
o
C. 
SAMPLE TEMP 
WARNING 
The Sample Temperature is less than 10
o
or is greater than 50
o
C. 
V/F NOT 
INSTALLED 
The CPU is unable to communicate with 
the V/F Board. 
Check and re-seat CPU and V/F 
board.  See Section 9.6.3. 
SYSTEM RESET 
A power Off-On cycle has occurred. 
None required. 
RAM  
INITIALIZED 
Dynamic memory has been re-initialized 
in response to the installation of a new 
PROM or memory chip. 
None required. 
O3 FLOW 
WARNING 
IZS flow rate is less than 1.8 L/M or >5.5 
L/M. 
Adjust pressure regulator to set 
proper flow rate. 