Photometric Calibrator Instruction Manual, 01124, Rev. J2
Table 9-1: Warning Messages
Warning Message
Meaning
Corrective Action
ANA LAMP
WARNING
WARNING
The O
3
REF value is greater than 5000 mV
or less than 2500 mV.
ANA LAMP
SHUTDOWN
SHUTDOWN
Temperature control of the Source Lamp
cannot be maintained at its 52
cannot be maintained at its 52
o
C set point.
O3 GEN LAMP
WARNING
WARNING
The Ozone Generator is unable to produce
at least 1000 ppb at its maximum output or
the O
at least 1000 ppb at its maximum output or
the O
3
GEN feedback control is adjusting
the Ozone generator drive signal by more
than a factor of two.
than a factor of two.
Check and adjust the O
3
GEN
lamp and reference det as
described in Section 9.6.8.
described in Section 9.6.8.
O3 GEN LAMP
SHUTDOWN
SHUTDOWN
Temperature control of the O
3
GEN Lamp
cannot be maintained at its 48
o
C
set point.
Check O
3
GEN lamp heater and
thermistor as described in
Section 8.10.
Section 8.10.
SAMPLE
PRESSURE
PRESSURE
The Sample Pressure is less than 15"Hg or
is greater than 35"Hg.
is greater than 35"Hg.
SAMPLE FLOW
WARNING
WARNING
The sample flow is less than 500 cc/min or
greater than 1000 cc/min. Check for
pneumatic system problems as described
in Section 9.6.1.
greater than 1000 cc/min. Check for
pneumatic system problems as described
in Section 9.6.1.
BOX TEMP
WARNING
WARNING
The inside chassis temp is less than 5
o
C or
is greater than 50
o
C.
SAMPLE TEMP
WARNING
WARNING
The Sample Temperature is less than 10
o
C
or is greater than 50
o
C.
V/F NOT
INSTALLED
INSTALLED
The CPU is unable to communicate with
the V/F Board.
the V/F Board.
Check and re-seat CPU and V/F
board. See Section 9.6.3.
board. See Section 9.6.3.
SYSTEM RESET
A power Off-On cycle has occurred.
None required.
RAM
INITIALIZED
INITIALIZED
Dynamic memory has been re-initialized
in response to the installation of a new
PROM or memory chip.
in response to the installation of a new
PROM or memory chip.
None required.
O3 FLOW
WARNING
WARNING
IZS flow rate is less than 1.8 L/M or >5.5
L/M.
L/M.
Adjust pressure regulator to set
proper flow rate.
proper flow rate.
9-4