Agilent Technologies 83491 Benutzerhandbuch

Seite von 60
2-9
Operation
To Compensate for Module Insertion Loss
To Compensate for Module Insertion Loss
The following steps allow you to enter an offset to compensate for the inser-
tion loss of the clock recovery module. This provides accurate amplitude mea-
surements at the input to the clock recovery module.
1
Disconnect the cable from the clock recovery module’s 
Input connector.
2
Measure the signal using a power meter. You can use either the 
Agilent 83480A’s built-in power meter or an external power meter.
3
Reconnect the cable to the clock recovery module.
4
Disconnect the cable from the reference receiver module’s input connector.
5
Measure the signal using a power meter. You can use either the 
Agilent 83480A’s built-in power meter or an external power meter.
6
Subtract the two measurements to determine the insertion loss of the module.
Insertion loss:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  ____________  
7
On the reference receiver module, press the front-panel channel 
SETUP key.
8
Press 
External scale, and set the Atten units to “decibel”.
9
Press 
Attenuation, and enter the value calculated i
Step 6
.