Operation
Using Probes with an Agilent 83491A
To compensate for a passive probe
1
Connect the probe to the 
Input connector on the Agilent 83491A clock recovery 
module. 
2
Attach the probe tip to the 
CAL hook that is located near the floppy disk drive. 
3
Press the reference receiver module’s front-panel channel 
SETUP key. 
4
Press 
Calibrate and then Calibrate probe
To compensate an Agilent 54701A active probe
1
Connect the Agilent 83491A output to the electrical measurement channel 
input.
2
Connect the probe to the 
Input connector on the Agilent 83491A clock recovery 
module. 
3
Connect the probe power cable to the 
Probe Power connector on the reference 
receiver module.
4
Attach the probe tip to the 
CAL hook that is located near the floppy disk drive. 
5
Press the reference receiver module’s front-panel channel 
SETUP key. 
6
Press 
Calibrate and then Calibrate probe