Toshiba LF410/LF602 User Manual

Page of 16
LF410/LF600  LF410/LF602 
3
 
totalizer. 
Fixed-value outputs — Outputs fixed-values for 
current and pulse outputs. 
Zero adjustment — Executes zero adjustment 
(on-stream at zero flow rate). 
Output signals 
Current output: 
4–20mAdc (load resistance  0 to 750
Ω) 
Note: The current output cannot be used with 
the PROFIBUS-PA ccommunication. 
Digital outputs —  One point (std.) and one more 
point is optionally available as follows. 
Digital output DO1 (std.): 
Output type: Transistor open collector 
Number of outputs: One point 
Output capacity: 30Vdc, 200mA maximum 
Digital output DO2 (opt.): 
Output type: Solidstate relay output (non 
polarity) 
Number of outputs: One point 
Output capacity: 150Vdc, 150mA maximum 
or 150 V ac (peak to peak), 100mA maximum 
DO1 and DO2 functions — One of the following 
functions can be assigned to DO1 (std.) and/or 
DO2 (opt.) 
• Pulse output (available only for DO1,DO2) 
Pulse rate:  3.6 to 36,000,000 pulses/hr (DO1) 
                     3.6 to      360,000 pulses/hr (DO2) 
(Over 3,600,000 pulses/hr, auto-setting) 
Pulse width: 0.5 to 500ms (but less than half of 
the period for 100% flow rate) 
Note: The same and simultaneous  pulse is not 
available between DO1 and DO2.) 
• Multi-range selection outputs (Note 1) 
• High, High high, Low, and/or Low low alarm 
outputs (Note 2) 
• Empty pipe alarm output 
• Digital Output Active Status (DO1 and DO2) 
(Note 2) 
• Preset count output 
• Converter failure alarm output 
Note 1: Two outputs (DO1 and DO2) are needed for 
4-range switching and forward/reverse 
2-range switching. 
Note 2: Normal Open (default set) or Normal Close 
is selected for alarm outputs when 
programming. 
The Status when power failure is kept to 
Normal Open. 
Communications output : 
• HART(std.) 
Digital signal is superimposed on 4–20mAdc 
current signal as follows: 
• Conforms to HART protocol 
Load resistance: 240 to 750
Ω 
Load capacitance: 0.25µF maximum 
Load inductance: 4mH maximum 
• PROFIBUS (opt.) 
Protocol : PROFIBUS-PA 
Baurate : 31.25kbps 
Bus voltage : 9-30Vdc 
Consumption electric current of bus:less than 
16mA 
Manufacture Ident-No. : 093B
HEX 
Standard Ident-No. : 9740
HEX
 
Slave address : 0-126 (Default address is 126) 
Profile : Profile Ver.3.01 for Process Control 
Devices 
Function blocks : AI(Flow)×1 , Totalizer×1 
LCD display: Full dot-matrix 128×128 dot LCD 
display (back-light provided) 
The data on the LCD inside the converter can 
rotate to 90, 180, and 270 degrees by a software, 
without rotating the indicator itself. (Combined 
type only) 
Parameter settings — Parameters can be set as 
follows:  
• IR Switches: Three key switches are provided to 
  set configuration parameters. 
• Digital communication: The AF900 hand-held 
 terminal or PROFIBUS is needed to set 
parameters. 
•Zero adjustment: Zero point adjustment can be 
started by pressing the switch in the converter. 
Damping: 0.5 to 60 seconds (selectable in 1 second 
increments) 
Zero and span calibration: Built-in calibration 
signal source allows converter unit check. 
Conditions when power fails: The outputs and 
display will remain as follows when power fails. 
Parameter setting values are stored in non-volatile 
memory and the values will be restored when the 
power returns to normal condition. 
• Current output: 0mAdc 
• Digital output:  OFF 
• LCD display:  No display 
• PROFIBUS:   No communication 
Power supply: One of the following can be selected: 
• 100 to 240Vac, 50/60Hz (std.) 
(allowable voltage 80 to 264Vac) 
• 24Vdc (allowable voltage 18 to 36Vdc) 
• 110Vdc (allowable voltage 90 to 130Vdc) 
Surge protection: Arresters are installed in the 
power supply, and current signal output circuit. 
Case: Aluminum alloy (equivalent of IP 67)  
Coating: Acrylic resin-baked coating, pearl-gray 
colored