Emerson 2200S User Manual

Page of 158
42
Micro Motion
®
 Model 2200S Transmitters
Configuring Process Measurement
3. If you do not know the standard density of your gas:
a.
Click the 
Gas Wizard
 button.
b.
Click 
Choose Gas
 and check the list. 
c.
If your gas is listed, select it and click 
Next
.
d. If your gas is not listed, click
 Enter Other Gas Property
 and provide the required 
information. You can describe your gas by molecular weight, its specific gravity, or 
density. If you use density, you must enter the density value in the configured density units 
and you must provide the temperature and pressure at which the density value was 
determined. When you are finished, click 
Next
.
e.
Verify the reference temperature and reference pressure. If these are not appropriate for 
your application, click 
Change Reference Conditions
 and enter new values for reference 
temperature and reference pressure.
f.
Click 
Next
. The calculated standard density value is displayed. 
If the value is correct, click 
Finish
. The value will be written to transmitter 
configuration.
If the value is not correct, click 
Back
 and modify input values as required.
Note: The Gas Wizard displays density, temperature, and pressure in the configured units. If required, 
you can configure the transmitter to use different units.
4.6.1
Volume or GSV flow cutoff
If the volume or GSV flow measurement drops below the configured volume or GSV flow cutoff, the 
transmitter will report a value of 0 for volume flow or GSV flow.
Note the following:
If Volume Flow Type is set to Liquid:
The volume flow rate is affected by the density cutoff. Accordingly, if the density drops 
below its configured cutoff value, the volume flow rate will go to zero even if flow is 
present and the volume flow rate is above the volume flow cutoff.
The volume flow rate is not affected by the mass flow cutoff.
If Volume Flow Type is set to GSV, the volume flow rate is not affected by either the density 
cutoff or the mass flow cutoff.
The volume or GSV flow rate reported via the mA output may also be limited by the AO 
cutoff. See Section 6.2.3 for more information.
Configuration
To configure the volume flow cutoff or GSV flow cutoff:
Using the Communicator, select 
Detailed Setup > Config Fld Dev Var > Flow
.
Using ProLink II, click 
ProLink > Configuration > Flow
.
Note: You cannot configure the volume flow cutoff or GSV flow cutoff with the display.