Emerson 2200S User Manual

Page of 158
44
Micro Motion
®
 Model 2200S Transmitters
Configuring Process Measurement
4.7
Configuring density measurement
The density measurement parameters control how the flowmeter measures and reports density. You 
must configure:
Density measurement units
Density cutoff
Density damping
Slug flow parameters
Note: If you use the display, you can configure only the density measurement unit.
4.7.1
Density measurement units
The default density measurement unit is g/cm3. See Table 4-6 for a complete list of density 
measurement units.
Configuration
To configure the density measurement unit:
Using the Communicator, select 
Detailed Setup > Config Fld Dev Var > Density
.
Using ProLink II, click 
ProLink > Configuration > Density
.
Using the display, see Figure 2-13.
4.7.2
Density cutoff
If the density measurement drops below the configured density cutoff, the transmitter will report a 
value of 0 for density.
Table 4-6
Density measurement units
Density unit
Unit description
Display
Communicator
ProLink II
SGU
SGU
SGU
Specific gravity unit (not temperature- 
corrected)
G/CM3
g/Cucm
g/cm3
Grams per cubic centimeter
G/L
g/L
g/l
Grams per liter
G/ML
g/mL
g/ml
Grams per milliliter
KG/L
kg/L
kg/l
Kilograms per liter
KG/M3
kg/Cum
kg/m3
Kilograms per cubic meter
LB/GAL
lb/gal
lbs/Usgal
Pounds per U.S. gallon
LB/CUF
lb/Cuft
lbs/ft3
Pounds per cubic foot
LB/CUI
lb/CuIn
lbs/in3
Pounds per cubic inch
ST/CUY
STon/Cuyd
sT/yd3
Short ton per cubic yard