Specifications
Specification 
¢
Dual-wavelength pulse type (665nm/880nm) arterial blood oxygen satura-
tion measuring instrument 
•  Name: 
Oxygen Saturation Monitor PULSOX-300i
¢
Functions:
•  Measuring range:  SpO
2
:  
0 to 100% SpO
2
   
Pulse rate:   30 to 230 bpm
•  Accuracy: 
SpO
2
±2% SpO
(70 to 100% range, 1S.D.)
   
Pulse rate:   ±2 bpm (30 to 100 range)
   
 
or ±2% of value  (100 to 230 range)
 
✻ 
PULSOX-300i has been clinically validated for use on adults.  
¢
Display
Display type: Liquid crystal display (with back-light function)
  Oxygen saturation (SpO
2
)
  Pulse rate number
  Pulse level meter (8 steps)
  Battery indication (3 steps)
  Error messages indication
¢
Warning functions
  SpO
2
: Low SpO
2
 warning (flashing display)
  Battery indication
  Error message indication
¢
Memory function
Up to 300 hours at one-second intervals
  SpO
2
 value, pulse rate, and error message information, and the file infor-
mation (file number, date and time of starting measurements, and duration 
of measurement)
   ✻ Data can be downloaded or displayed using the optional software.
¢
Operating conditions
  Temperature/humidity range
 
 
0 to 40°C (32 to 104°F); 30 to 85% relative humidity with no condensation
  Atmospheric pressure/altitude range
 
 
700 to 1060hPa (altitude: –400 to 3000m)
¢
Storage conditions
  Temperature/humidity range
 
 
–10 to 60°C (14 to 140°F); 10 to 95% relative humidity with no condensation
  Atmospheric pressure/altitude range
 
 
700 to 1060hPa (altitude: –400 to 3000m)
¢
Power
1 AAA-size battery: 1.5V   100mW (Service life under continual usage: 
approx. 30 hours with alkaline battery)
¢
Dimensions (W × H × D)
  68 × 58 × 15 mm