Calculation Method for Displayed Values
<Pulse level meter>
The pulse level meter displays the transmitted light amount ratio calculated by the 
following equation. The pulse level is indicated in eight levels (0 to 15%, full scale 
if over 10%).
<Oxygen saturation (SpO
2
)>
Updates the moving average of the oxygen saturation over the last 12 seconds at 
one-second intervals.
<Pulse rate (P.R.)>
Updates the moving average of the last 8 pulse rates at one-second intervals.
Pulse level (%) = ------------------------------------------------------------------------------------------------------------------ x 100
Transmitted light amount that does not vary due to the pulse
Transmitted light amount that varies due to the pulse