 Data example
 
Weighing data “kg” (+) 
S T
,
+ 0 0 1 2
.
3 4 5 _ k g
C
R
L
F
 
Weighing data “g” (-)   
S T
,
- 0 0 0 0 1 2 3 4 _ _ g
C
R
L
F
 
Counting data “pcs” (+) 
Q T
,
+ 0 0 0 1 2 3 4 5 _ P C
C
R
L
F
 
Out of weighing range (+)  O L
,
+ 9 9 9 9
.
9 9 9 _ k g
C
R
L
F
 
Data output mode (
prt
)
 
‰
 Command mode   
The scale is controlled by commands that come from an external device such as a 
personal computer. For details, refer to “13.2.3. Command mode”.   
‰
 Stream mode (
prt 0
Data is sent continuously. The data update rate is approximately 10 times per 
second, the same as the display refresh rate. 
There will be no output during the setting procedures. 
‰
 Print switch mode (
prt 2
When the weight display is stable, data is sent by pressing the 
 
PRINT
 
 switch.  
At this time, the display flashes once to indicate that the data is sent. 
‰
 Auto-print mode +/- data (
prt 3
When the weight display is stable at ±5d (d = minimum display in kg)
 
and above 
+5d or below -5d, the data is sent. The next transmission can not occur until after 
the weight display falls between –4d and +4d. 
‰
 Auto-print mode + data (
prt 4
When the weight display is stable at +5d (d = minimum display in kg)
 
and above, 
the data is sent. The next transmission can not occur until after the weight display 
falls +4d or below. 
 
Baud Rate (
bp5
)
 
Select the baud rate according to the device to be connected. 
‰
  2400 bps (
bp5 0
)
 
Select 2400 bps to connect to an AD-8121 printer. 
‰
  4800 bps (
bp5 1
‰
  9600 bps (
bp5 2